TREAT U BIOPHOTONS DEVICE Magnetic Field Safety Assessment

慈佑生命光子機磁場範圍安全性分析機制圖 Cropped

The actual test results show extremely low magnetic field strength, meeting international safety standards.

After actual measurement and analysis using a professional electromagnetic field measuring instrument (Lutron EMF-827 three-axis magnetic field sensor), the maximum magnetic field strength measured on the surface of the TREAT U BIOPHOTONS DEVICE is:

2.96 mG(milligauss)

This value is significantly lower than relevant international and domestic safety standards:

Safety standardsLimit value
ICNIRP4 mG
IEEE C95.6-20024 mG
China National Standard GB8702-20145 mG
Taiwan NCC Recommended Values10 mG

Actual measured maximum value is only:

2.96 mG 74%, compliant with the international standard of 4 mG

Therefore, it meets international safety regulations.


The magnetic field decays rapidly as the distance increases.

Actual test data shows:

DistanceMagnetic field strength
Host surface2.96 mG
5 cm0.73 mG
10 cm0.35 mG
15 cm0.20 mG
20 cm0.12 mG
25 cm0.08 mG
30 cm0.05 mG

It can be seen that the magnetic field strength decreases rapidly with increasing distance.

Distance from host:

10 When it is above 2 centimeters, the magnetic field is only 0.35 mG.

Approximately the international safety standard (4 mG) of:

1/11 Below

The magnetic field energy that users are exposed to in normal use environments is very low.


Safety margin analysis

0–10 cm

The magnetic field in the equipment's near-field operation area is still far below international safety standards.

4–8 inches

The magnetic field is significantly attenuated, belonging to a low magnetic field environment.

8–12 inches

The magnetic field is close to the environmental background value, making it safer.

30 cmAbove

The magnetic field is close to the natural background magnetic field in general living environments.


Comprehensive Security Conclusion

According to actual measurement data analysis:

✅ Maximum magnetic field value of only 2.96 mG

✅ Below relevant safety standards such as ICNIRP, IEEE, GB8702, and NCC

The magnetic field decays rapidly with distance.

✅ Over 10 cm is far below the international safety standard

✅ Normal operation will not cause excessive electromagnetic field exposure

Conclusion

The TREAT U BIOPHOTONS DEVICE has been verified through actual testing, and its magnetic field strength is far below the relevant international and domestic safety standards. Under normal operating conditions, it falls within the safe range and is safe to use. It can be confidently applied to daily health management and wellness maintenance environments.